Valve Drive ModuleThis schematic also available as:valve_drive.cdr (v.9), valve_drive.pdf Components
Click Functional DescriptionThis module supplies 120VAC power to the solenoid and directional valves used to direct fluid flow through the hydraulic system. Each valve-drive section (3 in all) supplies power as determined by its own selector switch. The user may manually turn a valve on/off or put it under electronic control via the "Ext. Ctrl. In" BNC connector. In this latter case, a TTL signal may control the valve drive; a high level will turn the drive on. The pilot light will illuminate when the valve is being powered (see Note 1 below). This module interfaces with the hydraulic system as follows: For Valve Drive section 1:
For Valve Drive section 2:
For Valve Drive section 3:
Note:The solid state relays within the module require a minimum load to shut off. If, for example, no valves are connected to the module then powering it on will cause all of the pilot lights to illuminate regardless of the position of the selector switches. In this case, however, the light intensity is significantly less than when a valve is connected and is turned on.
Last modified: February 18, 2002 Laura Galiana |